Siemens SC 7000/9000 five parameter patients monitor, with Masimo SPO2 module with all new accessories. CIE 1019

The Siemens SC 7000/9000 Patient Monitoring System can display various vital sign wave forms including 5-lead ECG, oxygen saturation (SP02/pulse oximetry), respiration rate, as well as blood pressure. This working monitor can be used as a “live” patient/actor hookup, displaying actual human vital signs in real time. This particular model includes a built-in DEMO mode that fills the screen with a variety of waveforms and vitals

Each unit includes:

5-lead ECG cables
Pulse Ox (SP02) Cable w/ Finger Probe
Blood Pressure Cuff w/ Tubing
Red Dot ECG Electrode Stickers (5-pack)
(Rolling stand not available for this monitor)
Siemens SC 9000XL/7000 Patients Monitor

DESCRIPTION:

Features:
Color display
Parameters: EGC, Resp, SpO2, NIBP, Temperature

SPECIFICATIONS:
Physical Weight & Dimensions
Height: 10 inches (254 mm)
Width: 14 inches (355.6 mm)
Depth: 4.5 inches (114.4 mm)
Weight: 18.2 lbs (8.2 kg)

ECG:
Display: up to 12 leads.
Measuring range: 15 – 300 1/min.
Accuracy:±2 1/min or±1% (whichever is greater).

Respiration:
Sensing leads: I or II (user-selectable).
Measuring method: impedance pneumography.
Measuring range: 0-155 breaths per minute.
Measuring accuracy: ±1 1/min or 2% of rate (whichever is greater)

SpO2:
Parameter display: saturation (%SpO2), pulse rate.
Measuring method: absorption – spectrophotometry.
Measuring range: SpO2: 1-100%.
Pulse Rate: 15-300 1/min.
Accuracy: SpO2 70-100%: ±2%, 0-69%: not specified.
Pulse rate: ±5 1/min or ±5% (whichever is greater)

NIBP:
Parameter display: systolic, diastolic, mean.
Measuring method: oscillometric technique.
Modes of operation: manual (single measurement), continuous (5 minutes) or interval.
Heart rate measuring range: 40-240 1/min.

Temperature:
Parameter display: absolute temperature, delta temperature.
Measuring range:
Absolute: -5°C to 50°C (23°F to 104°F)
Delta: 0°C to 55°C (32°F to 131°F)
Resolution: 0.1°C

Accuracy:
Absolute: ±0.1°C
Delta: ±0.2°C.