The Dash patient monitor can monitor many different parameters at the same time, some of these include 3 to 5 lead ECG, Pulse Oximetry, non-invasive blood pressure, temperature, and cardiac output. The Dash 3000 has an 8.4-inch color LCD screen where you can view 3 or 6 waveforms and battery capacity on the screen all at once.
The GE Dash 3000 comes with two batteries that can provide up to 5 hours of running time and can be recharged within 2 to 4 hours. The batteries can be charged one at a time as to lose any functionality of the monitor while charging.
Features:
- Networking options for both wired and wireless
- 3 to 6 waveform and up to 4 pressure can be views at once
- 4-5 hours of running battery time
- Multi-lead ECG analysis and DINAMAP NIBP technology all in one monitor
- Seamless Patient Data transfer to Facilities computer system
GE Dash 3000 Specifications:
Dimensions:
Height: 10.23 in (26 cm)
Depth: 7.8 in (20 cm)
Width: 11 in (28 cm)
Weight: 12 lbs (5.5 kg)
Display:
Size: 22 cm (diagonal).
Type: active-matrix color TFT.
Resolution: 640 x 480 pixels.
Number of traces: 6 maximum.
Number of seconds/trace: 4.9 at 25 mm/sec.
Sweep Speed: 6.25, 12.5, 25 mm/sec (with erase bar).
Information window: displays non-real-time information without obstructing the display of real-time information.
Display organization: prioritized by parameter.
Controls:
Trim Knob® control.
Five hard keys: Silence Alarm, Graph Go/Stop, NIBP Go/Stop, function (zero all) and Power On/Off.
Alarms:
Categories: patient status and system status.
Priority: 4-levels crisis, warning, advisory, message.
Notification: audible and visual.
Settings: default or user-defined.
Silencing: 1 minute, current alarm only.
Pause: 5 minutes in adult ICU mode, 3 minutes in Neonatal ICU mode, and 5, 15, minutes or permanent pause in OR mode.
ECG:
Standard leads available: I, II, III, V, aVR, aVL, and aVF.
Leads analyzed simultaneously: I, II, III and V (multi-lead).
Lead fail: identifies failed lead.
Alarms: user-selectable upper and lower heart rate limits.
Respiration:
Measurement technique: impedance variation detection.
Range respiration rate: 0-200 breaths per minute.
User-selectable upper and lower respiration rate limits, and user-selectable apnea limit.
Temperature:
Number of channels: 2.
Resolution: ±0.1°C.
Alarms: user-selectable upper and lower limits for T1, T2.
Invasive Blood Pressure
Number of channels: 2.
Range: -25mmHg to 300mmHg.
Alarms: user-selectable upper and lower limits for systolic, diastolic, and mean pressures.
NIBP:
Measurement technique: oscillometric.
Displayed parameters: systolic, diastolic, and mean pressures, time of last measurement.
Measurement modes: manual, auto, and stat in Adult ICU and OR modes, manual and auto in neonatal mode.
Systolic range: 30-275 mmHg adult, 30-235 mmHg pediatric, 30-135 mmHg neonate.
Diastolic range: 10-220mmHg adult, 10-220mmHg pediatric, 10-110 mmHg neonate.
Pulse Oximetry:
Parameters monitored: SpO2 and peripheral pulse rate.
SpO2 range: 50-100%.
PPR range: 30-300 beats per minute.
SpO2 accuracy: ±2%, 70 – 100% SpO2, ±3% (50-69% SpO2).
PPR Accuracy: ±3 beats per minute.