DETAILS:
Refurbished Siemens 300 & 300A Servo Ventilators offer a wide range of ventilation modes together with an Automode function for greater treatment possibilities, shorter weaning times, and increased patient comfort.
Features:
- Neonatal to adult patient range
- Wide range of ventilation modes
- Controlled, Support, or Combined ventilation modes
- Easy to use
- Optional graphical display, and compressor
- Refurbished
Configurable Model:
- Siemens 300A Servo Ventilator – #40300A
- Automode for fully adaptive patient interaction
- Neonate to adult patient range
- Wide range of ventilation modes: Controlled, Supported, Combined
- Easy to use
- Refurbished
Available Accessories For Product:
- Carts
- Breathing Systems
- Compressors
- External Power Supply
- Humidifiers
Specifications:
Dimensions:
- Patient unit: 9.5″ (W) x 14.6″ (D) x 9.4″ (H)
User interface: 17.0″ (W) x 15.9″ (D) x 12.8″ (H) - Weight: Approx. 52.9 lbs
Ventilation Modes:
- Controlled
- Supported
- Combined
Patient Range:
- Large/Medium/Small Animal
Classification:
- IEC60601-1/EN60 601-1 Type B
Method of Triggering:
- Flow and Pressure
- Flow Range: 0.1 ml/s – 3 l/s
- Power Supply: 100, 120, 220, and 240 VAC 10%, 50-60Hz
Alarms:
- Airway Pressure (Upper): 15-120 cm H2O
- High Continuous Pressure: Set PEEP-level +15 cm H2O for more than 15 seconds.
- O2 concentration: Set value 6%. Lower alarm limit cannot go below 18% or exceed 90%.
- Expired Minute Volume (Upper Alarm Limit): Adult/Pediatric Range: 0-60 l/min, Neonate Range: 0-6 l/min.
- Expired Minute Volume (Lower Alarm Limit): Adult/Pediatric Range: 0.3-40 l/min, Neonate Range: 0.06-4 l/min.
- Apnea: Adult: 20 s. Pediatric: 15 s. Neonates: 10 s.
Monitoring:
- Frequency: CMV or SIMV frequency 1% or 0.5 b/min.
- Pressures: Peak, Pause, Mean and End exp. 5% b/min or 2 cm H2O
- Insp. Tidal Volume: Adult Range:50-3999 ml 6% or 3 ml. Pediatric Range: 10-399 ml 6% or 0.5 ml. Neonate Range: 2.0-39 ml 6% or 0.5 ml.
- Exp. Tidal Volume: Adult Range: 50-3999 ml 6% or 3 ml. Pediatric Range: 10-399 ml 6% or 0.5 ml. Neonate Range: 2.0-39 ml 6% or 0.5 ml.
- Exp. Minute Volume: Adult Range: 4.0-60 l/min 6% or 0.1 l/min. Pediatric Range: 1.0-5.0 l/min 6% or 0.05 l/min. Neonate Range: 0.20-1.50 l/min 6% or +/-0.02 l/min.
- Flow Rate: 0.00-3.00 l/s or 0.00-180 l/min 1% or 1% or 1 digit.
- O2 Concentration: 5% of read value
- Ventilation modes: Controlled, Supported, Combined.
- Patient range: Adult, Pediatric, Neonate.
- Classification: IEC60601 – 1/EN60 601-1 Type B.
- Dimensions: Patient unit: W 242 x D 370 x H 240 mm.
- User interface: W 431 x D 150 x H 325mm.
- Weight: Approx. 24kg.
- Method of triggering: Flow and Pressure.
Reviews