The Oxylog 3000 is a non-invasive ventilation machine with advanced leak composition. the system can be used in a hospital setting, in the field for emergencies, and can be used in winged aircraft for both adult and pediatric patients.
Features:
- High confidence during transport.
- Wide range of ventilation modes and AutoFlow.
- Integrated capnography.
- Ready for pediatric patients.
- Automatic altitude compensation (BTPS).
- Intuitive user interface.
- Advanced data export functionality.
- Drager Oxylog 3000 Specifications
Specifications:
Dimensions:
- Height: 7.2 In (18.4 cm)
- Width: 11.4 In (29 cm)
- Depth: 6.9 In (17.5 cm)
- Weight (Including internal battery): 12.8 lbs (5.8 kg).
Display:
- Technology: Electro-luminescence (EL).
- Resolution: 240 x 128
- Dimensions: 4.3 x 2.2 In (10.8 x 5.6 cm).
Wireless Network:
- Standards: IEEE 802.11g, Wi-Fi compatible.
- Frequency Range: 2.412 to 2.462GHz.
Ventilation Modes:
- VC-CMV, VC-AC, VC-SIMV, SpnCPAP, PC-BIPAP.
Ventilation Settings:
- Pressure support: in the ventilation modes VCSIMV, PCBIPAP and SpnCPAP.
- Apnoea ventilation: in the ventilation mode SpnCPAP.
- AutoFlow (optional): in the ventilation modes VC-CMV, VCAC and VC-SIMV.
- NIV: in the ventilation modes: SpnCPAP (/PS), PC-BIPAP (/PS), VC-CMV /AF, VC-AC /AF and VC-SIMV/AF.
- Ventilation Respiratory Rate.
- 2 to 60 /min (VC-SIMV, PC-BIPAP).
- 5 to 60 /min (VC-CMV, VC-AC).
- 12 to 60 /min for apnoea ventilation.
General:
- Tidal volume: VT 0.05 to 2.0 L; BTPS.
- Ti / I:E: I:E or Ti configurable, for all ventilation modes.
- Ventilation Time Tatio I:E: 1:100 to 50:1.
- Inspiration Time Ti: 0.2 to 10 s.
- Inspiratory Pressure Pinsp: PEEP +3 to +55 mbar.
- O2 Concentration: 40 to 100 Vol.%.
- PEEP / CPAP: 0 to 20 mbar.
- Trigger Sensitivity (Flow Trigger): 1 to 15 L/min.
- Pressures support Psupp: 0 to 35 mbar (relative to PEEP)
- Slope (Pressure Rise Time): Slow, standard, fast.